46
c)
Do not use hot air sterilization and radiation sterilization as this may
result in damage to the product.
d)
Please use the recommended sterilization procedures for
sterilization. It is not recommended to sterilize with sterilization
procedures such as ethylene oxide, formaldehyde and low temperature
plasma sterilization. The manufacturer assumes no responsibility for the
procedures that have not been recommended. If you use the sterilization
procedures that have not been recommended, please adhere to related
effective standards and verify the suitability and effectiveness.
* Fractional pre-vacuum procedure = steam sterilization with repetitive
pre-vacuum. The procedure used here is to perform steam sterilization
through three pre-vacuums.
6.4.6
Storage
1.
Store in a clean, dry, ventilated, non-corrosive atmosphere with a
relative humidity of 10% to 93%, an atmospheric pressure of 70KPa to
106KPa, and a temperature of -20 °C to +55 °C.
2.
After sterilization, the product should be packaged in a medical
sterilization bag or a clean sealing container and stored in a special
storage cabinet. The storage time should not exceed 7 days. If it is
exceeded, it should be reprocessed before use.
Notes:
a)
The storage environment should be clean and must be disinfected
regularly.
b)
Product storage must be batched and marked and recorded.
6.4.7
Transportation
1.
Prevent excessive shock and vibration during transportation, and
handle with care.
2.
It should not be mixed with dangerous goods during transportation.
3.
Avoid exposure to sun or rain or snow during transportation.
7
Storage, maintenance and transportation
7.1
Storage
7.1.1 This equipment should be stored in a room where the relative
humidity is 10% ~ 93%, atmospheric pressure is 70kPa to106kPa, and the
temperature is -20°C ~ +55°C.
7.1.2 Avoid storage in too hot conditions. The high temperature will
shorten the life of electronic components, damage batteries, reshape or
melt some plastic.