8
2.2.1.3
CCW Mode
The motor handpiece rotates in a counterclockwise direction only. This
mode is used to inject calcium hydroxide or any other intracanal
medicament. When this mode is used, a double-beep sound will be heard
continuously.
2.2.1.4
REC Mode
Reciprocating mode.
F: Forward angle, R: Reverse angle
It is suggested that the difference between the forward angle and
reverse angle should be greater than or equal to 120 degrees, otherwise,
root canals cannot be prepared effectively.
Forward Angle<Reverse Angle, such as F:30/R:150, the effective
cutting angle is Reverse Angle, it is suitable for use of the reciprocating
files such as WOODPECKER W3-ONE.
Forward Angle>Reverse Angle, such as F:180/R:30, the effective
cutting angle is Forward Angle, it is suitable for use of the reciprocating
files such as SENDONELINE S1.
2.2.1.5
ATR Mode
ATR: Adaptive Torque Reverse function.