24
After selecting the file system, press the
Setting button “P” to enter the select file
number, press the adjusting button “+”/“-” to
select the file number. Press the main button to
confirm.
The parameters of "W3-Pro" can also be
changed from the default setting.
To return to the default setting, press the
Setting button “P” for a long time to enter the
preset program during standby state, select
"W3-Pro" and press the "main" button to
confirm, the default setting will be reloaded,
Turn off the motor handpiece and subsequently
turn on the power to restore the default setting.
Changing the preset program default setting
is not recommended to prevent file
separation.
4.4
Handpiece functions setting
4.5
With the motor handpiece turned off, hold down the Setting button “P”
and press the main button to enter handpiece functions setting, press
Setting button “P” till target setting, press the Adjusting button “+”/“-” to
adjust, then press the main button to confirm.
With the motor handpiece turned off when the
Setting button “P” held down and pressing the
main button to enter the handpiece functions
setting, the software version number will appear
on the display screen.
After 3 seconds of displaying the version
number on the screen, the "Auto Power OFF
“time setting can be changed by, pressing the
adjusting button “+”/“-” and then press the
"Main" button to confirm.
If no buttons are pressed, the auto power-off
time
of motor, the handpiece can be set from 3 to
30 minutes in 1-minute increments
Summary of Contents for Ai-Motor MotoPex
Page 4: ...2 ...
Page 5: ...3 ...
Page 42: ...40 the contra angle ...
Page 45: ...43 6 4 4 Drying If the cleaning and disinfection process does not have an automatic ...
Page 56: ...ZMN SM 034 V1 9 20210430 ...