E32.0.02.6C-04
Manual for connection and operation of the
GIA 2000 / AAG and GIR 2002 / AAG
page 29 of 30
Display range:
(voltage-, current and frequency-measurement)
-1999 ... 9999 digits, initial value, terminal value and decimal point position arbitrary.
Recommended range: < 2000 Digit
Accuracy:
(at nominal temperature)
Standard-signals:
< 0.2% FS ±1Digit (from 0 – 50mV: < 0.3% FS ±1Digit)
RTD:
< 0.5% FS ±1Digit
Thermocouples:
< 0.3% FS ±1Digit (from Type S: < 0.5% FS ±1Digit)
Frequency:
< 0.2% FS ±1Digit
Point of compensation
:
±1°C ±1Digit
(at nominal temperature)
Temperature drift:
< 0.01% FS / K (from Pt100 – 0.1°C: < 0.015% FS / K)
Measuring freq.:
approx. 100 measures / sec. (standard-signal) or
approx. 4 measures / sec. (temperature-measurement) or
approx. 4 measures / sec. (frequency, rpm at f > 4 Hz) or accordingly f (at f < 4 Hz)
Display:
approx. 13 mm height, 4-digit red LED-display
Operating:
4 push-buttons or via interface
Interface:
EASY
BUS
-interface, electrically isolated
Transmitter supply:
24 V
DC
±5%, 20mA, electrically isolated
or the corresponding designation on the label on the housing
Outputs:
1 volt-free Relay-outputs
*
2
(standard)
or the corresponding designation on the label on the housing
1 analog output, electrically isolated
Output 1 *
2
:
change-over contact, breaking capacity: 10A (ohmic load), 250 V
AC
Response Time:
< 25 msec.
for standard signals
< 0.5 sec.
for temperature, frequency (f > 4 Hz)
Output-functions:
2-point, min-/max-alarm
Switching points:
arbitrary
Analog output:
0-20 mA (AAG420), 4-20 mA (AAG420) or 0-10V (AAG010)
or the corresponding designation on the label on the housing
Scaling:
arbitrary
Accuracy:
0.3 % FS
Min. permissible load
:
1000 Ohm (at AAG010)
Max. perm. burden
:
300 Ohm (at AAG420, AAG020)
Power supply:
230 V
AC
, 50/60 Hz (standard)
or the corresponding designation on the label on the housing
Nominal temp.:
25°C
Operating ambient:
-20 to +50°C
Relative humidity:
0 to 80% RH (non condensing)
Storage temp.:
-30 to +70°C
Housing:
Dimensions:
48 x 96 mm (front-panel dimensions).
Installation depth:
approx. 115 mm (incl. screw-in/plug-in clamps)
Panel Mounting:
with brackets
Panel cut-out:
43.0
+0.5
x 90.5
+0.5
mm (H x W)
Connection:
via screw-in/plug-in clamps
Conductor cross-selection from 0.14 to 1.5 mm²
Protection class:
front IP54, with optional mounting seals IP65
EMC:
EA1 +A2 (appendix A, class B), additional errors: < 1% FS
When connecting long leads adequate measures against voltage surges have to be taken.
*
2
= not available at GIA 2000
14. Disposal notes
This device must not be disposed as ‘residual waste’.
To dispose this device, please send it directly to us (adequately stamped). We will dispose it appropriately
and environmentally friendly.