T2 Elite 2/Pro 2/Express 2/Elite/Pro/Express Service Manual
87
Storage problems
The message "Media disks getting full..."
appears.
The media disks are
reaching maximum
capacity.
Delete unused clips and empty
the Recycled Bin to make
space on the media drives.
"Rebuild" appears in "Status" on the BIOS
screen during T2 startup.
Playback/Record stops. The performance is
low.
The icon (
) is displayed at the
right of the screen for the Front Panel Mode
or on the status bar of the Workstation
Mode.
Running RAID
re-building.
The RAID rebuild
function is automatically
executed if the T2 comes
to an abnormal end (such
as if the main power fails
or if you hold down the
power switch to terminate
the T2 forcefully) when
used the T2 last time.
Wait until the RAID rebuild
completes (until 100% appears
on the screen).
"Degraded" or "Failed" appears in "Status"
on the BIOS screen during T2 startup.
The icon (
) appears at the right of
the screen for the Front Panel Mode or on
the status bar of the Workstation Mode.
The T2 starts up in the maintenance mode
even when starting up in the normal mode.
RAID system has some
problem.
Verify the RAID system,
replace faulty storage unit, and
build/rebuild a RAID volume.
Refer to
and rebuilding a RAID
volume" on page 44
.
Problem
Possible Causes
Corrective Actions