Handbuch OSMOMAT 050
Seite 51 von 51
Version 1.1 (2011-10-12)
4. Screw the counter nut against the measuring cell to secure the system.
5. Assembly of the lower measuring cell
Replace the large O ring (63.22 x 1.78 mm) in the lower half of the cell and reinstall the cell
half into the housing from below using the two screws.
Insert the yellow LED into the designated bore and secure using cable tie.
Plug the connector of the pressure transducer into the mainboard and replace the lower hous-
ing cover using the six screws.
The OSMOMAT 050 is ready for use again.