OSMOMAT 050 User Guide
Page 27 of 51
Version 1.1 (2011-10-12)
After a few seconds, the display shows "00.0". The baseline of the measurement system for the value
"00.0" is set with the fill level taken into account. The waste bottle must be placed on the bottle tray
and the tube system has to be filled without air bubbles.
3. Move the bottle from the tray to the table surface. The height difference between tray and ta-
ble surface is exactly 10 cm. Setting a stable measurement value of approx. 10 cm water col-
umn should not take longer than 10 seconds.
Figure 5-4: Setting the value for 10 cm water column
4. Adjusting the potentiometer will set the measurement value to exactly 10.00 cm water column.
Figure 5-5: Adjustment to 10 cm water column