While the device is not in use, the measuring cell may dry out, especially in the area of
the pressure transducer. The salt crystals that form adhere to the pressure measure-
ment membrane and can destroy it. Regular rinsing prevents the drying out.
Make sure to top off the Ringer's solution and empty the waste bottle regularly.
If the device is left in standby mode for an extended period of time without topping of
the Ringer's solution, the measuring cell will dry out quickly, which may result in the
destruction of the pressure measurement membrane.
If the device is not used for an extended period of time (several weeks), the measuring
cell must be opened and rinsed using distilled water. Afterwards, the device can be
switched off.
When When working on the pressure transducer, the device must always be switched on.