Seite
50
von
51
Handbuch OSMOMAT 050
Version 1.1 (2011-10-12)
1. If you already have a new lower measuring cell, you have to remove the old O- ring.
2. Clean the lower measuring cell with cleaning solution and rinse it with water very carefully.
Install the new O ring (2 mm) in the lower measuring cell.
3. Screw the pressure transducer into the lower half of the measuring cell. It may be necessary
to back off the counter nut a bit. Screw in the pressure transducer with moderate force until
the O ring is pressed against the cell and the ring becomes visibly deep dark.