Handbuch OSMOMAT 050
Version 1.1 (2011-10-12)
13 Service Instructions for Replacing a Defective
Pressure Transducer
13.1 When is a replacement of the pressure transducer required?
The pressure transducer must be replaced in the following cases:
1. The OSMOMAT 050 no longer switches to STAND BY mode
2. A continuous alarm sounds, even when the measuring cell is open
3. The solvent added to the opened measuring cell flows downward through the pressure
transducer
The pressure transducer can be replaced by trained medical technicians by following these instruc-
tions.
A new pressure transducer can be ordered using order no. 50.3.0010. The serial number of your de-
vice must be communicated because devices up to serial number 101122 also require replacement of
the lower measuring cell due to a component change by the manufacturer of the sensor.
Package contents:
•
1 pressure transducer, complete with board and cable
•
1 O ring
•
1 cable tie