Components 

Function 

Malfunction 

Possible Cause 

Measure 

 

The red "Overload" 
LED lights up (when 
injecting a sample or 
during automatic rins-
ing of the measuring 
cell using solvent) 

The system is partially clogged between 
the measuring cell and the waste bottle 

All tubes behind the measur-
ing cell, the plunger of the 
waste bottle, the channel and 
the outlet cannula in the upper 
half of the measuring cell must 
be cleaned 

 

Liquid leaks from the 
tube connection be-
tween measuring cell 
and pump when in-
jecting the sample 

Entry of measuring cell is clogged 

Open and clean measuring 
cell 

Pressure  
transducer 

Device rinses the 
measuring cell in pe-
riodic intervals and no 
longer switches to 
STAND-BY position 
Device finds no base-
line characteristic 

Pressure transducer is defective (pressure 
measurement membrane destroyed) 
 
The measuring cell was dry for an extend-
ed period of time and salt crystals have 
formed 

Pressure transducer must be 
replaced 
 
Open measuring cell and rinse 
using distilled water 

 

Solvent leaks from the 
septum during injec-
tion. Salt crust forms 
below the septum 
screw 

Septum leaking 

Replace septum 

 

Summary of Contents for OSMOMAT 050

Page 1: ...www gonotec com User Guide Colloid Osmometer Oncometer OSMOMAT 050 Osmolality Measurement in medical and pharmaceutical field...

Page 2: ......

Page 3: ...marked as follows This symbol warns of the danger of corrupt ing measurement results for example by improper use of the measuring cell This symbol warns of the danger of damag ing the unit or the meas...

Page 4: ...050 Colloid Osmometer 9 1 6 Packaging Contents 9 2 1 Overview of Display and Connectors 10 2 2 Power Supply 12 2 3 Dimensions and Weight 12 2 4 Safety and Handling Information 12 2 5 Known Risks Assoc...

Page 5: ...cks 6 of Medical Devices Operator Ordinance 38 8 3 Measurement Checks 11 of Medical Devices Operator Ordinance 38 8 4 Internal Quality Control of the OSMOMAT 050 39 8 5 Replacing a Defective Power Fus...

Page 6: ...tions The OSMOMAT 050 colloid osmometer determines the colloid osmotic pressure COP of blood plasma The main diagnostic considerations for determining the colloid osmotic pres sure in medical applicat...

Page 7: ...ressure e g serum or plasma to the sample cell permeation of solvent causes underpressure in the reference cell based on the laws of osmosis because the membrane is impenetrable to proteins this can o...

Page 8: ...ide Version 1 1 2011 10 12 1 4 Reproducibility of the OSMOMAT 050 Measurement display 3 digits Measuring range 73 mm HG 100 cm water column or 9 99 kPa Resolution 0 1 mm HG over the entire measurement...

Page 9: ...the contents of your shipment are complete We cannot accept responsibility for any missing items reported at a later date Accessories and consumables included in the shipment Item number 20 9 0100 1 p...

Page 10: ...AND BY OSMOMAT 050 Colloid Osmometer 23 2 Connection tube between supply bottle and upper half of measuring cell Bottle tray Supply bottle Waste bottle Connection tube between waste bottle and ground...

Page 11: ...power supply Characteristics for device fuse to be used Manual pump control RS232 output Analog output Alignment potentiometer for analog port Serial number plate with information on serial number pow...

Page 12: ...e performed Always switch on device when performing work on the measuring cell or the pressure trans ducer see chapter 3 Open measuring cell and remove membrane and filter paper Repeatedly rinse measu...

Page 13: ...switched on and in calibration mode can be set by pressing the CALI BRATION button Excess pressure which may result in the destruction of the pressure transducer is indicated on the display This happ...

Page 14: ...s solution and empty the waste bottle regularly If the device is left in standby mode for an extended period of time without topping of the Ringer s solution the measuring cell will dry out quickly w...

Page 15: ...re 4 1 Making the tube connections Figure 4 1 Making the tube connections Use only the Tygon tubes supplied by the manufacturer Other tubing material is per meable to air which causes air bubbles that...

Page 16: ...nd waste bottles are passed through a bore of the unscrewed cap and pushed onto the corresponding plungers Figure 4 2 The plungers hold the tubes on the bottom of the bottles The second opening on the...

Page 17: ...contain any amount of solved gas air If necessary it must first be degassed by heating up just under the boiling point or by degassing under vacuum We recommend using the 30 mL plastic syringe includ...

Page 18: ...pore size or pore size range is the typical factor for the retention and is called cut off The mem branes used for membrane osmometry usually have a cut off between 5 000 and 30 000 dalton The pore si...

Page 19: ...rom a few seconds to an hour Next the membrane is completely submerged into the water and ready for installation into the mem brane osmometer for measurement of aqueous solvents It is important to ins...

Page 20: ...use in the measuring cell Degassing the filter paper requires a lot of care because all air bubbles must be removed from the paper It is recommended to degase use to degas the filter paper together wi...

Page 21: ...sounds stop unscrewing immediately After pressure com pensation the alarm is silenced and you can proceed to open carefully Loosen the tension screw flip back the clamping yoke and remove the upper h...

Page 22: ...r bubbles form Next remove the water from the measuring cell using a paper tissue Figure 4 8 Adding the degassed solvent The lower half of the measuring cell is now cleaned and you can add the solvent...

Page 23: ...f the measuring cell without air bubbles Remove the conditioned semipermea ble membrane from the water flip it once and place it on top of the filter paper in the measuring cell with the active side f...

Page 24: ...ed together carefully in small intervals If the pressure increases to above 99 9 cm water column the display starts flashing Figure 4 12 If the pressure increases further to approx 550 cm water column...

Page 25: ...semipermeable membrane becomes slightly transparent and a uniform somewhat dark ring is visible at the edge of the membrane Figure 4 14 The displaced Ringer s solution collected in the overflow chann...

Page 26: ...performed directly by creating a hydrostatic pressure difference To fill all tubes without air bubbles press the Pumpe button at the back panel until the tube going to the waste bottle is also filled...

Page 27: ...tube system has to be filled without air bubbles 3 Move the bottle from the tray to the table surface The height difference between tray and ta ble surface is exactly 10 cm Setting a stable measuremen...

Page 28: ...es for Measurement The OSMOMAT 050 prepared according to item 4 and calibrated according to item 5 as well as a precision syringe with special cannula for injecting the sample included with the access...

Page 29: ...ssible without significant changes to the colloid osmotic pressure How ever the sample should not be frozen 2 The minimum sample volume required is 150 L The samples must be homogeneous and should hav...

Page 30: ...cannula through the septum as far as it will go It is impossible to puncture the membrane during this procedure Inject the sample into the measuring cell using three equal cycles of approx 50 L Fig 6...

Page 31: ...ize in the lower measuring cell filled with solvent This balanced state is detected by the electronics and the measurement value is shown on the digital display Figure 6 6 At the same time the LED 2 a...

Page 32: ...ristic see Page 34 Figure 6 9 After the RESULT LED and LED 3 light up the OSMOMAT 050 waits again for the same time elapsed from injection until display of the first measurement value to accept and st...

Page 33: ...g to curve 2 If obtaining the result takes a very long time due to a sluggish adjustment characteristic curve 3 Fig ure 6 9 the measurement value is accepted after 4 5 minutes and the second measureme...

Page 34: ...value 1st measurement value 2nd measurement value 2nd measurement value 1st measurement value 2nd measurement value Curve 3 Adjustment characteristic of the osmotic pressure when the pores of the semi...

Page 35: ...oups identifying the function of each component its potential malfunctions the effects of the malfunctions on the measurement sys tem the possible causes of the malfunctions and the procedure for corr...

Page 36: ...value is unsatisfactory mea surement value too low Pump tube too old flagged Part of the injected sample flows toward the pump System partially clogged between measur ing cell and waste bottle Strong...

Page 37: ...aks from the tube connection be tween measuring cell and pump when in jecting the sample Entry of measuring cell is clogged Open and clean measuring cell Pressure transducer Device rinses the measurin...

Page 38: ...inance 8 2 Safety Checks 6 of Medical Devices Operator Ordinance The OSMOMAT 050 is not listed in Annex 1 of the Medical Devices Operator Ordinance Public au thorities do not require the unit to under...

Page 39: ...ctive power fuse unplug the unit from the power supply Risk of electric shock To replace the fuses use a small screwdriver to remove the fuse holder on the rear of the unit The two fuses can now be re...

Page 40: ...gram is sent Idle level 3V U 7 5V Active level 3V U 7 5V DSR Dataset Ready input This signal can prevent transmission of a telegram The OSMOMAT 050 starts transmitting only after DSR becomes active Th...

Page 41: ...with special cannula for septum injection 1 50 9 0130 Plastic bottles 250 mL supply waste bottles 2 50 9 0140 Tool for widening the tube ends 1 50 9 0150 Tweezers for membranes 1 50 9 0160 Plunger for...

Page 42: ...atement OSMOMAT 050 Gonotec GmbH GSG Hof Reuchlinstr 10 11 D 10553 Berlin We hereby declare that the OSMOMAT 050 colloid osmometer complies with Directive 98 79 EEC Compliance is declared per Annex II...

Page 43: ...nal warranty services such as on site service may be available from an authorized Gonotec sales partner B Limitation of the Warranty TO THE EXTENT PERMISSIBLE UNDER THE APPLICABLE LOCAL LAWS NEITHER G...

Page 44: ...returning the equipment 4 If the equipment is being returned for credit you must include all accessories power cord software disks manuals etc 5 Return the equipment in its original packaging If you...

Page 45: ...ibration direct pressure calibration through hydrostatic solvent col umn or using solutions of known concentration Measurement display 3 digit Recorder output 2 mV digit adjustable 10 for strip chart...

Page 46: ...pen 3 The solvent added to the opened measuring cell flows downward through the pressure transducer The pressure transducer can be replaced by trained medical technicians by following these instruc ti...

Page 47: ...50 Seite 47 von 51 Version 1 1 2011 10 12 13 2 Instructions 13 2 1 Disassembly 1 Unscrew and remove lower front panel 6 screws 2 Unplug pressure transducer connector from board 3 Release LED cable by...

Page 48: ...housing 5 Remove the lower measuring cell and the large O ring 63 22 x 1 78 mm 6 Loosen the counter nut and unscrew and remove the pressure transducer from the measuring cell Remove the small O ring...

Page 49: ...er is delivered with a cable tie and a new O ring For devices up to serial number 101122 where the pressure transducer is replaced for the first time the lower measuring cell must be replaced as well...

Page 50: ...h cleaning solution and rinse it with water very carefully Install the new O ring 2 mm in the lower measuring cell 3 Screw the pressure transducer into the lower half of the measuring cell It may be n...

Page 51: ...e the large O ring 63 22 x 1 78 mm in the lower half of the cell and reinstall the cell half into the housing from below using the two screws Insert the yellow LED into the designated bore and secure...