OSMOMAT 050 User Guide
Page 33 of 51
Version 1.1 (2011-10-12)
Pressing the “RESULT CHECK”- button will display the second measurement value.
If the 2nd measurement value deviates from the first value by more than two digits, LED 6 lights up
(Figure 6-8)
, lower molecular parts permeate.
Figure 6-8: Display with adjustment characteristic according to curve 2
If obtaining the result takes a very long time due to a sluggish adjustment characteristic
(curve 3, Fig-
ure 6.9)
, the measurement value is accepted after 4.5 minutes and the second measurement value is
stored after an additional 4.5 minutes, LED 4 lights up.
After storing the second measurement value, five rinse cycles using solvent are performed in the
measuring cell. During rinsing and zero point setting, LED 7 lights up. After zero point setting is com-
plete, LED 8 lights up. A shift of the zero point is indicated by LED 9.