Page 32 of 51
OSMOMAT 050 User Guide
Version 1.1 (2011-10-12)
6.5 Recording and Storing the Second Measurement Value
An automatic measurement cycle consists of recording and storing the measurement result, storing a
second measurement result, rinsing the measuring cell, and checking the zero point.
The functions in addition to storing and displaying the first measurement value allow the evaluation of
the quality of the measurement result in connection with the adjustment characteristic
(see Page 34,
Figure 6.9)
. After the RESULT LED and LED 3 light up, the OSMOMAT 050 waits again for the same
time elapsed from injection until display of the first measurement value to accept and store the second
measurement value.
After the second measurement value is stored, LED 4, 5, or 6 light up. The measurement value is
stored until a new measurement is performed. Afterwards, the measuring cell is automatically rinsed
and LED 7 lights up.
A comparison of both measurement values shows the adjustment characteristic.
If the adjustment characteristic matches the
curve shape 1 in Figure 6.9
, LED 5 lights up.
Figure 6-7: Display with adjustment characteristic according to curve 1