To adjust the rotation of the horizontal arm and
vertical arm, work on the four nuts placed on the bent
tube, rotating them with equal intensity
Rotate clockwise to have greater frictional force and
vice versa to loosen the force for rotating the arm.
5.2 Extraordinary maintenance
For extraordinary maintenance of the lamp the following need to be checked:
•
luminous intensity reduced by more than 30%: check that an opaque mist has not formed on the
internal surface of the glass protection with the evaporation of detergents penetrated inside the
head and evaporated due to the heat of the bulb.
•
detachment of parts of the paint from the apparatus due to accidental collision with other objects:
these fragments can fall onto the patient zone during the use of the device.
For the cases mentioned above, contact the Manufacturer or the assistance centre nearest the Client
authorised by the Manufacturer.
5.3 Solving faults and anomalies
n Problem
Solution
1
The lamp does not hold its position
Check the horizontally of the plate installed on the wall
and the adjustment of the frictional force
2
The lamp does not function
Check the presence of the fuses inside the terminal
board and their state. Check the connection of the
electrical connectors, the presence of bulbs, their
condition and the presence of tension at the bulb holder.
3
The fuse continues to burn
Check the characteristics of the inserted fuses
(paragraph 4.4)
+
-
Summary of Contents for SATURNO WALL
Page 30: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 30 of 34 6 2 Explode drawing...
Page 31: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 31 of 34...
Page 32: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 32 of 34...
Page 33: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 33 of 34...
Page 34: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 34 of 34...