MO003-GB 18/05/06
Operative Manual
Rev.0
Page 23 of 34
The handle should be sterilised every time necessary and in any case following eventual internal protocols
or national or local directives.
,
The handles that can be sterilised are subject to natural wear. Normally they last about
200 cleaning cycles. Damaged handles can no longer be used. The user must take
account of the demands of the competent national body on hygiene and disinfection.
The GIMA Company does not advise hot air sterilisation. If, however, it becomes necessary, the handle
must be sterilised, unpacked, at 134°C for 3 minutes.
Handle disposal:
sterilise the now worn-out handle to avoid contamination and eliminate it according to the normal
procedures for non biologically infected materials.
4.3 Changing the bulb
Symbol on the lamp support that indicates “Danger of burning” (do not touch the quartz
stem of the halogen bulb with bare hands, use a cloth or some soft paper)
Turn the illumination system off and let it cool down (danger of
burning), positioning the switch (1) of the command panel (2) of the
lamp on 0, off.
Use both the hands for rotate anti-clockwise sense the porthole with
attack to bayonet. Therefore making allows the complete uncoupling of
the protection glass and the access to the optical zone.
1
2
Summary of Contents for SATURNO WALL
Page 30: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 30 of 34 6 2 Explode drawing...
Page 31: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 31 of 34...
Page 32: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 32 of 34...
Page 33: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 33 of 34...
Page 34: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 34 of 34...