MO003-GB 18/05/06
Operative Manual
Rev.0
Page 28 of 34
4
The lamp flickers
The central rotor charcoal does not have good surface
contact on the electrical tracks. Open the inspection
cover and check the condition visually.
Clean the tracks and position the charcoal so that it
adheres perfectly to the collector.
The electronic board or dimmer is faulty
5
The luminous band on the operating
field is not in focus
Check the correct insertion of the halogen bulb (it must
be perpendicular to the bulb holder)
6
The bulb breakage frequently
With a voltmeter verify the tension to clips 3 and 4. (to
see prescription to paragraph 3,4).
Verify the presence of tension peaks on the net.
Probable that is caused from the ignition of other
systems, which to es. conditioned air.
5.4 Ordinary spare parts list
Description
Ordering code
75W 12V Halogen bulb
AL75/12
Handle that can be sterilised
Z142166