MO003-GB 18/05/06
Operative Manual
Rev.0
Page 24 of 34
Pay attention in the uncoupling of the porthole because there is no
coupling with head.
N.B: closing the porthole make attention to insert the three springs of
support in the 3 appropriate rabbet and rotate clockwise until the lock of
detention.
Remove the burnt-out halogen bulb and insert a new GIMA 75W 12V
original halogen bulb in the holder.
Attention!
Do not hold the bulb of the halogen lamp with your bare hands.
,
Only use original GIMA halogen bulbs.
If other bulbs are used, the entire guarantee on the maintenance of the technical data of
the lamp and the perfect functioning of the product (luminous intensity and focus) is
forfeited.
4.4 Changing fuses
,
Before proceeding to the operations of changing fuses, remove the supply to the electric
panel from the main line switch dedicated to the scialytic lamp
The fuses are placed on the cover of the slab. Unscrew the fuse and replace the out of order fuse with one
new of identical characteristics, being respected the face on the label.
Data on the fuses established by GIMA:
n° terminal
Description
Type of fuse
Dimensions
1
On the line of the input phase
T1A
5 x 20
2
On the line of the neutral input
T1A
5 x 20
4
On the line of the output phase
T6,3A
5 x 20
Summary of Contents for SATURNO WALL
Page 30: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 30 of 34 6 2 Explode drawing...
Page 31: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 31 of 34...
Page 32: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 32 of 34...
Page 33: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 33 of 34...
Page 34: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 34 of 34...