MO003-GB 18/05/06
Operative Manual
Rev.0
Page 14 of 34
2.1.1 Plate anchorage with wall in CONCRETE (reinforced concrete)
Mechanical anchorage:
fix the plate with n° 4 Hilti HSL-G-TZ M12/25 mod. expansion dowels,
scrupulously following the indications issued by the Hilti manufacturing company and reported here for
information:
Anchorage stays
d
(mm)
do
(mm)
t
(mm)
hs
(mm)
l
(mm)
Mt
(Nm)
SW
(mm)
x
(mm)
HSL-G-TZ M12/25
12 18 105 93 125 80 19 25
do
Nominal point diameter
Mt Moment of tightening twist
t
Minimum perforation depth
Sw Opening the wrench
hs
Minimum insertion depth
x
Fixing
height
l
Length of anchorage stays
1.
Mark all 4 fixing holes at the chosen point in the ceiling.
2.
Make the first hole in conformity to the diameter of the safety
anchorage.
3. With a pump or a vacuum cleaner with a tube terminal remove
the dust and small perforation fragments from the hole.
4. Make the anchorage stay enter the hole holding it along the axis
and using a hammer.
Attention!
Keep the depth of the insertion in mind.