3.2 Installation oscillating arm
Unscrew the three screws TSCE M4x8 from the circular metal ring of
the oscillating arm (1).
Join the electrical connectors (2) respecting the colours of the electrical
cables:
•
yellow/green with yellow/green
•
blue with blue
•
brown with brown
Align the hinge (1) of the vertical oscillating arm with the horizontal
arm to (2). Keep aligned on the same axis the parts to couple and align
the hole of the contact crawling (3) with the threaded hole of the
horizontal arm (4) that will have then to be centred from screw TSCE
M3x15.
Insert the hinge of the oscillating arm in the horizontal structure being
lend cure to the operation, avoiding to force the union if there is not
right position in the connection.
Push until end of vertical oscillating arm in the horizontal arm.
Prepare three screws TSCE M4x8 with a veil of Loctite 243 on
everybody and screwing them to the horizontal arm holes.