GIMA SpA
MT BASE PLUS
MA321IGBa
Manuale d’Istruzioni/Instruction’s Manual
37
75 - 99 Hz by the suitable keys. The + key is used to increase the frequency value, while the –
to decrease it.
D
ISPLAY
I
NTENSITY
L
EVEL
(4)
On the central part of the unit, at the right, there is a display to visualize the intensity values, in
Gauss. Switching on the unit, the intensity is at 30 Gauss and can be changed (from 10 to 80
Gauss, with 5 steps) by the suitable keys. The + key is used to increase the intensity value,
while the – to decrease it.
D
ISPLAY
T
IME
T
IMER
(5)
At the left of the unit, there is a display to visualize the time values, in minutes. Switching on
the unit, the time is at 30 minutes and can be changed (from 0 to 60 min) by the suitable keys,
when the apparatus does not supply. The + key is used to increase the frequency value, while
the – to decrease it. By the start key
(5c)
it is possible to start the time count and start the
output too. The output start condition is indicated by the lighting of a light point at right of
minutes’ display. By the stop key
(5d),
it is possible to stop the time count and stop the output
too. The stop condition is indicated by the switching off of the point at the right of the minutes’
display. The end of the treatment time is indicated by an acoustic signal of 10 seconds. By
setting the 00 time it’s possible to realize an infinite treatment time, without programmed
interruption.
C
ONNECTOR
OUT
1
(6)
Connection point of the supplied magnetic transducer. If there is current output , the
corresponding warning light, warning light 1, is lighted (patient connected).
C
ONNECTOR
OUT
2
(7)
Connection point of the supplied magnetic transducer. If there is current output , the
corresponding warning light, warning light 2, is lighted (patient connected).