OPERATION AND
MAINTENANCE MANUAL
EN
MU_027
Rev.1
29/04/2020
Page 6 of 33
2.2
Safety conditions (secondary effects)
Optical safety
Possibility of glare.
-
Do not direct the light source into the patient’s and/or
operator’s eyes.
-
When Product use is restricted to the face (maxilla-facial
surgery, plastic surgery, ear-nose-throat surgery) the patient’s
eyes must be covered with adequate protection.
Failure to follow such precautions could cause glare and
potential damage to the retina.
Electromagnetic disturbance
To avoid any significant risk of reciprocal interference due to the
presence of the Product during specific exams or treatments, refer
to section 10 of the Manual.
Incorrect use
-
Never place and/or hang anything on the Product.
Failure to follow such precaution could result in such objects
falling in the operating area.
Do not place objects on Product.
-
Never hang on the Product with the body weight of a person.
Failure to follow such precaution could damage the Product
structure.
-
Never cover the head of the Product during operation to prevent
overheating.
-
Avoid the Product parts colliding with one another or other
nearby equipment.
Knocks could cause the detachment of plastic parts or paint from
the Product which could fall in the patient area.
Improper use of mobile version
Pushing or resting on the Product
is forbidden.
In the case of the mobile version, do not rest, push or lie on the
product. Failure to comply could result in damage to the product
and to devices nearby and injury to staff members.
2.3
Environmental conditions
-
The Product is not suitable for use in explosion-risk areas.
-
The Product is not suitable for use wherever there are
inflammable mixes of anaesthetics with air, oxygen or N
2
O
(laughing gas).
-
The Product is not suitable for use in environments rich in
oxygen and use is not intended in the presence of inflammable
agents.
-
During operation, the ambient temperature must be between
10°C and 40°C.
-
Relative humidity must be between 30% and 75%.
-
Atmospheric pressure must be between 700 and 1060hPa.
Summary of Contents for GIMAled81
Page 37: ...MANUALE D USO E MANUTENZIONE IT MU_027 Rev 1 29 04 2020 Pag 13 di 33 Modello a piantana...
Page 57: ...MANUALE D USO E MANUTENZIONE IT MU_027 Rev 1 29 04 2020 Pag 33 di 33 Note...
Page 94: ...OPERATION AND MAINTENANCE MANUAL EN MU_027 Rev 1 29 04 2020 Page 13 of 33 Mobile model...
Page 114: ...OPERATION AND MAINTENANCE MANUAL EN MU_027 Rev 1 29 04 2020 Page 33 of 33 Notes...