OPERATION AND
MAINTENANCE MANUAL
EN
MU_027
Rev.1
29/04/2020
Page 20 of 33
7.3
Periodical checks to be performed on
the Product
Perform the Product electrical
check.
At the time of start up and after each maintenance job, perform
electrical tests and jobs indicated in the IEC 62353 standard.
Making any changes to this device
is forbidden.
7.4
Routine maintenance
Interrupt the power supply before
doing any maintenance jobs.
Check Product integrity.
N
Period
Action
1
Before using
Make sure there are no pieces or fragments of paint that could
become detached and fall within the operating field. If there are any,
remove them manually.
2 Before using
Make sure the light source protection screens are not damaged. If
they are, contact the Customer Service.
3 Once a year
Check all the Product joints and make sure there are no noises or
squeaks. If there are, lubricate the clutches involved with suitable
grease for industrial use at a service temperature between -30°C
and + 120°C, type OKS 470 or with similar properties.
4 Once a year
If the Product fails to maintain a regular position, adjust the clutches
as indicated at points
7.1 and 7.2 (arm and clutch adjustment)
.
5
Once a year
(CEILING VERSION)
Make sure the bar retention screws (1) are tightened properly. Also
check the bar horizontal arm retention screws (4). If these are not
properly fastened, adequately tighten.
To access the nuts (1), slide off the aluminium ring (2) downwards by
loosening the screws and bar cover (3).
Summary of Contents for GIMAled81
Page 37: ...MANUALE D USO E MANUTENZIONE IT MU_027 Rev 1 29 04 2020 Pag 13 di 33 Modello a piantana...
Page 57: ...MANUALE D USO E MANUTENZIONE IT MU_027 Rev 1 29 04 2020 Pag 33 di 33 Note...
Page 94: ...OPERATION AND MAINTENANCE MANUAL EN MU_027 Rev 1 29 04 2020 Page 13 of 33 Mobile model...
Page 114: ...OPERATION AND MAINTENANCE MANUAL EN MU_027 Rev 1 29 04 2020 Page 33 of 33 Notes...