MAINTENANCE 

 

The 

CORSIA

 atomiser does not need maintenance or lubrication. 

Before use always check correct functioning and safety of the device. Carry out disinfection as described in the “CLEANING 
ACCESSORIES” section. Unpack the instrument and 

always check 

integrity of plastic parts and feeding cable, they might have 

been damaged during previous use. 
Connect cable to electrical network and turn switch on. 
Close the air outlet with one finger to make sure that noise produced is regular and there is no malfunctioning.  
With the air outlet always closed check the correct functioning of the nebulization regulator by turning the knob form MIN to 
MAX.  Make sure the indicator of the pressure meter is working correctly. 
Verify that the atomiser is not damaged by previous use (it was badly put away or badly knocked).  
A protection fuse (

F 1x1.6A 250V

) reachable from exterior and it situated in the plug protects the instrument.  

For use replacing, always check the type and the range indicated. 

 

Fault type

 

Cause

 

Solution

 

1. The device doesn’t work 

a) The plug may be misplaced in the wall 

socket 

b) Thermal protector may be on (the device has 

been working beyond its limits and / or near 
heat sources) 

a)  Make sure the plug is properly placed in the wall 

socket. Make sure the ON/OFF switch is in position 

I.

 

b)  Switch off the device by pressing the switch to 

position 

and left the motor cool down for at least 30 

minutes. 

2. Low  Nebulization 
 

Clogged Nebulizer Tank 

Clean and disinfect the nebulizer tank as explained in the 
instruction manual 

3. Low  Nebulization 

 

Clogged Nebulizer Tank 

If cleaning was not succesful change cruet 

4. Absence of Nebulization 

 

Clogged Nebulizer Tank 
 
Air tube is bended or squeezed 

a)  Check that the nebulizer contains medication; Make 

sure that the nebulizer is not clogged;                          
Check the connection between the compressor air 
outlet port and the accessories 

b)  Make sure the air tube is not bended or squeezed. 

5. Slow  Nebulization 

 

Highly dense drug 

Dilute drug in physiological liquid 

6. Noisy  Device 

 

Extended use 

Call retainer or manufacturer GIMA S.p.A. 

Fault 1 - 2 - 3 - 4 - 5 - 6  

No solution with previous items

Call retainer or manufacturer GIMA S.p.A. 

 

If 

the unit does not nebulizer once the above conditions have been checked, we suggest to contact your dealer or technical 

service GIMA S.p.A. 
 

The manufacturer  will provide upon request electric diagrams, component list, descriptions, setting instructions and 
any other information that can help the technical assistance staff for product repair. 
 

 

 

 

BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT  

GIMA S.p.A. TECHNICAL SERVICE 

GIMA S.p.A. DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICL SERVICE CHECKING, APPEARS 

TO BE TAMPERED 

 

 
 

 

 
 
 

1- Air 

Tube 

2- Nebulizer 

Tank 

3- Nebulization 

Nozzle 

4- Nebulizer 

Top 

5- Mouthpiece 
6- Adult 

Mask 

7- Pediatric 

Mask 

8- 

Nosepiece (accessory under  
request) 

Summary of Contents for CORSIA

Page 1: ...il gima gimaitaly com www gimaitaly com INTERNATIONAL Tel 39 02 953854209 221 225 Fax 39 02 95380056 e mail export gimaitaly com www gimaitaly com AEROSOL CORSIA CORSIA AEROSOL EQUIPMENT CORSIA AEROSO...

Page 2: ...componenti originali forniti dal costruttore al fine di garantire la massima efficienza e sicurezza del dispositivo Non immergere mai l apparecchio in acqua Posizionare l apparecchio su superfici pia...

Page 3: ...sensi dell art 13 del Decreto Legislativo 25 Luglio 2005 n 151 Attuazione delle Direttive 2002 95 CE 2002 96 CE e 2003 108 CE relative alla riduzione dell uso di sostanze pericolose nelle apparecchiat...

Page 4: ...3 2 Classe A Fluttuazioni di tensione flicker IEC EN 61000 3 3 Conforme Guida e dichiarazione del costruttore Immunit Elettromagnetiche Test di Emissione Livello di prova Guida all ambiente elettroma...

Page 5: ...erchio porta filtro togliere il filtro e inserire quello di ricambio Inserire di nuovo il coperchio porta filtro nell apposita sede Premere l interruttore sulla posizione I per procedere con la nebuli...

Page 6: ...omponenti con acqua tiepida e o sapone neutro risciacquare abbondantemente rimuovendo l eccesso di acqua e fare asciugare all aria in luogo pulito ACCESSORI IN DOTAZIONE ACCESSORI Kit Accessori HI FLO...

Page 7: ...moprotettore il dispositivo ha lavorato al di fuori dei limiti di funzionamento prescritto e o ha lavorato vicino a fonti di calore o ambienti con temperature elevate Accertarsi che la spina sia ben i...

Page 8: ...GIMA giudicher l apparecchio non idoneo alla riparazione per visibili segni di contaminazione esterne e o interne render l apparecchio al cliente con chiaramente specificato APPARECCHIO NON RIPARATO a...

Page 9: ...he highest efficiency and safety of the device Never immerge the appliance into water Position the appliance on flat stable surfaces Position the device in a way that the air inlets on the back aren t...

Page 10: ...accessories with a cloth soaked in methylated spirits or hypochlorite based solutions Put the instrument and accessories in a bag with indication of disinfecting We also request to specify the kind o...

Page 11: ...oluzione NaCl 0 9 MMAD 3 25 m GSD 3 45 WEIGHT 2 20 Kg SIZE 255 x 190 x 165 h NOISE LEVEL measured as specifications of EN 3544 1 Approx 57dB A DUTY CYCLE to 40 C and 110 operating voltage Not Stop Ope...

Page 12: ...SORIES Procced as follows to disinfect the accessories 1 Turn the upper part of the nebulizer in an anti clockwise direction 2 Disconnect the internal pisper at the base of the nebulizer using the fin...

Page 13: ...ower supply Mains power quality should be that of a typical commercial environment or hospital Surge IEC EN 61000 4 5 1kV differential mode Mains power quality should be that of a typical commercial e...

Page 14: ...nt Open the air filter cover remove the filter and insert the new one Place back the cover Press the ON OFF switch to position I to start nebulization To interrupt or stop the treatment press again th...

Page 15: ...switch to position 0 and left the motor cool down for at least 30 minutes 2 Low Nebulization Clogged Nebulizer Tank Clean and disinfect the nebulizer tank as explained in the instruction manual 3 Low...

Page 16: ...ent Utiliser seulement les accessoires et les composants originaux Ne jamais plonger l appareil dans l eau Positionner l appareil sur des surfaces planes et stables Positionner l appareil de fa on vit...

Page 17: ...uer la r paration dans les plus brefs d lais Il est donc requis de lire attentivement le mode d emploi pour eviter d endommager l appareil a la suite d un usage impropre Veuillez toujours specifier le...

Page 18: ...UX TERMES DE LA DIRECTIVE EUROP ENNE 2002 96 EC Au terme de son utilisation le produit ne doit pas tre limin avec les d chets urbains Le produit doit tre remis l un des centres de collecte s lective p...

Page 19: ...tuels d p ts de salet Traiter le n buliseur et ses parties comme cela est indiqu ci dessous sauf le tuyau d alimentation de l air Au cas o l on effectuerail l application sur un autre patient ou au ca...

Page 20: ...est utilise dans un tel environnement Test d Emission Niveau de test Guide l environnement lectromagn tique D scharges lectrostatiques ESD IEC EN 61000 4 2 6kV en contact 8kV dans l air Les sols devra...

Page 21: ...hors des limites de fonctionnement prescrit et ou a travaill pr s de sources de chaleur ou environnements hautes temp ratures S assurer que la fiche est bien ins r e dans la prise de courant S assure...

Page 22: ...t ins rer le filtre de rechange Ins rer de nouveau le couvercle porte filtre dans son logement Appuyer sur interrupteur sur la position I pour proc der avec la pulv risation Pour suspendre ou terminer...

Page 23: ...rstellers um die h chste Leistungsf higkeit und Sicherheit zu gew hrleisten Tellen Sie das Inhalationsger t auf eine ebene und stabile Oberfl che Das Ger t muss so positioniert werden dass die Bel ftu...

Page 24: ...iner Hypochloritl sung getr nkt wurde Die Zubeh rteile sind in die gleiche Desinfektionsl sung zu tauchen Ger t und Zubeh r im desinfizierten Zustand in einem Beutel mit dieser Angabe stecken Beim Zus...

Page 25: ...odukt nicht zusammen mit dem Hausabfall beseitigt werden Es kann zu den von den st dtischen Beh rden eingerichteten Sammelstellen oder zu den Fachh ndlern die einen R cknahmeservice anbieten gebracht...

Page 26: ...sorgf ltig reinigen alle Arzneimittelreste und eventuelle Shmutzablagerungen entfernen Behandeln Sie das Spr hger t und alle Teile mit Ausnahme des Luftschlauchs entsprechend der folgenden Beschreibun...

Page 27: ...as Ger t unter diesen Bedingungen benutzt wird Emissionstests Pr fniveau Elektromagn Umgebung Elektrosatische Entladungen ESD IEC EN 61000 4 2 6kV bei Kontact 8kV in der Luft Die B den m ssen aus Holz...

Page 28: ...ters Entfernen Sie den Filter und setzen Sie einen neuen ein Abdeckung wieder schlie en Dr cken Sie den EIN AUS Schalter auf Stellung I damit wird die Zerst ubung eingeschaltet Um den Betrieb des Ger...

Page 29: ...th lt Vergewissern Sie sich das die Zerst uberd se nicht verstopft ist berpr fen Sie die Verbindung zwischen dem Anschluss der Luftaustritts ffnung des Kompressors und dem Zubeh r 4 Keine Zerst ubung...

Page 30: ...onentes originales Nunca sumergir el aparato en agua Posicionar el aparato sobre superficies planas y estables Posicionar el aparato de modo tal de evitar de ocluir las tomas de aire situadas en la pa...

Page 31: ...un pa o humedecido en alcohol desnaturalizado o soluciones a base de ipoclorito y los accesorios sumergi ndolos en las mismas soluciones desinfectantes Introducir en una bolsa specificando que se tra...

Page 32: ...Volumen m ximo de llenado 6ml Exactitud de la presi n metro 5 Condiciones de funcionamiento Temperatura ambiente 10 40 C Porcentaje de humedad ambiente 20 85 RH Altitud 0 2000m s l m Condiciones des c...

Page 33: ...eletr nicos Emisiones irradiadas conducidas CISPR11 Clase B El aerosol CORSIA es adecuado para ser usado en toods los ambientes incluso en aquellos dom sticos y en aquellos conectados directamente a...

Page 34: ...ecomendamos usar el nebulizador durante 6 meses o un m ximo de 120 tratamientos El nebulizador CORSIA debe ser substituido despu s de un per odo largo de inactividad si est deformado o roto o si el in...

Page 35: ...co puede estr en el dispositivo ha estrado trabajando por enzima de sus l mites y o cerca de fuentes de calor Aseg rese que el enchufe est correctamente colocado en la toma de corrente Aseg rese de qu...

Page 36: ...e de reemplazo abrir la cubierta del filtro de aire retire el filtro e insertar el nuevo Coloque de nuevo la cubierta Presione el interruptor ON OFF a la posici n I para iniciar la nebulizaci n Para i...

Reviews: