13
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The
CORSIA
aerosol is intended for use in the electromagnetic environment specified below.
The customers or the user of the
CORSIA
aerosol should assure that it’s used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
Irradiated / Conducted emissions CISPR11
Group 1
The CORSIA aerosol only used RF energy
only for its internal functioning.
Therefore its RF emissions are very low and
are not cause interference in
proximity of any Electronic appliances.
Irradiated / Conducted emissions CISPR11
Class [B]
The CORSIA aerosol can be used in all
environments, including domestic and
those connected directly to the public mains
distribution that supplies power
to environments used for domestic scopes.
Harmonic emissions
IEC/EN 61000-3-2
Class [A]
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Complies
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The
CORSIA
aerosol is intended for use in the electromagnetic environment specified below.
The customers or the user of the
CORSIA
aerosol should assure that it’s used in such an environment.
Immunity Test
Compliance
Electromagnetic environments - guidance
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
± 6kV on contact
± 8kV in air
Floors should be wood, conceret or
ceramic tile. If floors are coverei with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient / burst
IEC/EN 61000-4-4
± 2kV power supply
Mains power quality should be that of a
typical commercial environment or
hospital.
Surge
IEC/EN 61000-4-5
± 1kV differential mode
Mains power quality should be that of a
typical commercial environment or
hospital.
Loss of voltage, brief voltage interruptions
and variations
IEC/EN 61000-4-11
5%U
T
for 0.5 cycle
40%U
T
for 05 cycle
70%U
T
for 25 cycle
<5%U
T
for 5 sec
Mains power quality should be that of a
typical commercial environment or
hospital If the user of the CORSIA
aerosol request that the appliance
operates continuosly, the use of a
continuity unit is recommended.
Magnetic field
IEC/EN 61000-4-8
3A/m
The power frequency magnetic field
should be measured in the intended
installation location to assure that it’s
sufficiently low.
Conducted Immunity
IEC/EN 61000-4-6
3Vrms 150kHz to 80MHz
(for appliances that aren’t life - supporting)
-
Irradiated Conducted
IEC/EN 61000-4-3
3V/m 80MHz to 2.5 GHz
(for appliances that aren’t life - equipment)
-
Note U
T
is the value of the power supply voltage
STANDARD ACCESSORIES
ACCESSORIES
HI-FLO KIT
(Nebulizer HI-FLO, Adult Mask, Pediatric Mask,
Air Tube and Mouth-piece)
Air Filter (+n°3 spare part)
For each individual patient it’s recommended to use the nebulizer for 6 months or for a maximum of 120 treatments.
The nebulizer must be replaced after a long period of inactivity, if it is deformed or broken, or if the nebulizer nozzle is blocked
by dry medicine, dust, ecc..
Only use the original nebulizer supplied with the device
The air filter must be replaced every 25 hours of functioning or when it result particulary worn.
For replacement, lift the filter and replace with a new one.
Only use original filter
DON’T USE THE DEVICE WITHOUT AIR FILTER