Series 8500A Peak Power Meters
1-2
Manual 20790, Rev. C, November 1998
1.1.1
Environmental Requirements
Series 8500A power meters are type tested to MIL-T-28800E, Type III, Class 5, Style E for Navy
shipboard, submarine, and shore applications except as follows:
•
Operating temperature range is 0 °C to 50 °C (calibrator operating temperature range is 5 °C to
35 °C).
•
Non-operating (storage) temperature range is -4 0°C to +7 0°C.
•
Relative humidity is limited to 95% non-condensing.
•
Altitude and EMI requirements are not specified.
1.1.2
Items Furnished
In addition to options and/or accessories specifically ordered, items furnished with the instrument are:
1 ea. - Power Cord
1 ea. - Detachable Detector Cable (for Model 8501A), or
2 ea. - Detachable Detector Cables (for Model 8502A)
1 ea. - Operation and Maintenance Manual
1.1.3
Items Required
The 8500A requires an external power detector; see Appendix D for power detector specifications.
1.1.4
Tools and Test Equipment
No special tools are required to operate the Series 8500A. Test equipment required for calibration or
performance verification is described in Chapter 4.
1.1.5
Cooling
No cooling is required if the instrument is operated within its specified operating temperature range
(0 to 50 °C).
1.1.6
Cleaning
The front panel can be cleaned using a cloth dampened with a mild detergent; wipe off the detergent
residue with a damp cloth and dry with a dry cloth. Solvents and abrasive cleaners should not be used.
1.1.7
Receiving Inspection
Use care in removing the instrument from the carton and check immediately for physical damage, such
as bent or broken connectors on the front and rear panels, dents or scratches on the panels, broken
extractor handles, etc. Check the shipping carton for evidence of physical damage and immediately
report any damage to the shipping carrier.
Each Giga-tronics instrument must pass rigorous inspections and tests prior to shipment. Upon receipt,
its performance should be verified to ensure that operation has not been impaired during shipment. The
performance verification procedure is described in Chapter 5 of this manual.
Summary of Contents for 8501A
Page 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Page 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Page 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Page 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Page 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Page 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Page 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Page 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Page 262: ...Series 8500A Peak Power Meters Index 8 Manual No 20790 Rev C November 1998 ...