Decimal
Value
Condition
52
Calibration aborted for lack of a detector
53
Calibration passed, but with PROM read error
54
Calibration failed with PROM read error
55
Calibration failed
57
Autozero aborted for lack of detector
58
Autozero fail
59
Autoscaling aborted for lack of detector
60
Autoscaling interrupted by GPIB bus command, no trigger received
61
Unable to Autoscale
63
Unable to place markers specified or required to make a pulse parameter measurement
170
Self test failed
See Self-Test Failure Indications in Section 3.6 for a description and possible causes pertaining to Error Numbers 01
through 11. These will be decimal values 170 through 180, respectively
184
New software detected. Must recalibrate both detectors
185
Serial numbers don’t match. Must recalibrate A and B detectors.
186
Serial number doesn’t match. Must recalibrate detector A
187
Serial number doesn’t match. Must recalibrate detector B
189
Instrument operation failure. PPM will probably need to be reset
190
Under range. Measurement not accurate
191
Over range. Damage to instrument may have occurred
Series 8500A Peak Power Meters
3-42
Manual No. 20790, Rev C, November 1998
Summary of Contents for 8501A
Page 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Page 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Page 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Page 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Page 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Page 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Page 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Page 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Page 262: ...Series 8500A Peak Power Meters Index 8 Manual No 20790 Rev C November 1998 ...