7.
Press [F1].
8.
The DVM should read -5.000 V
±
35 mV.
9.
If the unit being tested is an 8502A, press [F2] and repeat Steps 3 through 8 after moving the
rear cable to the ANALOG B output.
E. Thermistor Sense Circuit.
1.
Press [CLEAR] [4] [F1]
2.
The reading should be approximately 5000 mV at a detector temperature of 30 °C (82.4 °F). The
reading will vary about
±
0.2 V/°C with any increase or decrease in temperature from the 30 °C
level (a new reading is taken each time F1 is pressed.)
3.
Press [CLEAR] [CLEAR].
Series 8500A Peak Power Meters
5-14
Manual No. 20790, Rev C, November 1998