5.3.3
A2 Regulator Board
A. +5 V Adjust
1.
Connect the DVM LO test lead to A2TP11.
2.
Connect the DVM HI test lead to A2TP9.
3.
Adjust A2R42 (+5V ADJ) for a DVM reading of +5.000V
±
100 mV.
4.
Vary the Variac voltage from 100 Vac to 130 Vac. The DVM reading must not change more
than
±
5 mV.
B. -5.2 V Check.
1.
Connect the DVM HI test lead to A2TP10. The DVM must read -5.200 V
±
200 mV.
2.
Vary the Variac voltage from 100 to 130 Vac. The DVM reading must not change by more
than
±
5.2 mV.
3.
Set the Variac to 115 Vac.
C. -15 V Adjust.
1.
Connect the DVM LO test lead to A2TP6.
2.
Connect the DVM HI test lead to A2TP5.
3.
Adjust A2R22 (-15 V ADJ) for a DVM reading of -15.000V
±
15 mV.
4.
Vary the Variac voltage from 100 to 130 Vac. The DVM reading must not change by more
than
±
15 mV.
D. +15 V Check.
1.
Connect the DVM HI test lead to A2TP3. The DVM must read +15.000 V
±
45 mV.
2.
Vary the Variac voltage from 100 to 130 Vac. The DVM reading must not change by more than
±
15 mV.
Turn OFF the power.
Disconnect the Variac and connect the power meter to the ac power source.
Testing and Calibration
Manual No. 20790, Rev C, November 1998
5-11
Summary of Contents for 8501A
Page 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Page 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Page 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Page 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Page 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Page 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Page 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Page 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Page 262: ...Series 8500A Peak Power Meters Index 8 Manual No 20790 Rev C November 1998 ...