5.2 Performance Testing
The tests in this section are to determine that the performance of the Model 8501A or 8502A Peak Power
Meters (PPM) is within the parameters of the specifications given for the instrument. Assemble all of the
required test equipment and the PPM to be tested. Connect all equipment to the ac power line, turn it on,
and allow at least 30 minutes for it to warm up before starting the test.
5.2.1
Equipment Required
The equipment listed in Table 5-1 is required for Performance Testing. Equivalent or similar test
equipment can be substituted.
Description
Representative Model
Key Characteristics
CW Thermistor Power Meter
HP Model 432B
Instrument accuracy of at least
0.5%
Thermistor Mount
HP Model 8478B
≤
1.1 SWR @ 1 GHz
Digital Voltmeter (DVM)
Fluke Model 8600A
±
0.05% accuracy
Directional Coupler
Narda Model 3002, 10 dB
≤
1.15 SWR @ 1 GHz
Step Attenuator, 0 to 50 dB in
10 dB increments
Weinschel Model 1
≤
1.15 SWR @ 1 GHz Accuracy
<0.1 dB
RF Source (Signal Generator)
Wavetek Model 2510A Option XP
>+20 dBm @ 1 GHz & pulse mod
Low Pass Filter
HP Model 360B
>30 dB attenuation @ 2 GHz
Precision Pulse Generator
Wavetek Model 178
10 ppm Frequency Accuracy
Return Loss Bridge
Wiltron 63N50 or 97NF50
Directivity 36 dB. Freq. Range
from 1 GHz to max. freq. of
detector in use
Scalar Network Analyzer
Giga-tronics Model 8003
Frequency Range as above
Precision DC Source
Datel Model DVC 8500A
0.05% Accuracy
Graphics Plotter
HP Colorpro Model 7440A
GPIB compatible
Table 5-1. Required Test Equipment
Series 8500A Peak Power Meters
5-2
Manual No. 20790, Rev C, November 1998
Summary of Contents for 8501A
Page 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Page 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Page 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Page 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Page 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Page 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Page 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Page 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Page 262: ...Series 8500A Peak Power Meters Index 8 Manual No 20790 Rev C November 1998 ...