4.2.7
Calibrator PC Assembly (A12)
Refer to Figure 4-9 and schematic diagram #17097 in Chapter 8.
The RF Diode Detector Automatic Calibrator uses a computer-controlled calibration circuit to generate
precise calibration data using a simple power meter (thermistor) as a reference. This permits diode
detectors to be used for absolute power measurements with the same accuracy usually associated with
thermistor power meters.
Since the automatic calibration system is a new concept developed by Giga-tronics for calibrating diode
detectors, the first portion of this discussion will be a general description of the overall theory involved
in the system operation. The last part of the discussion will cover the operation of the specific
components contained in the calibrator assembly.
Figure 4-9. Detector Calibrator (A12) Block Diagram
Theory of Operation
Manual No. 20790, Rev C, November 1998
4-27
Summary of Contents for 8501A
Page 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Page 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Page 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Page 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Page 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Page 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Page 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Page 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Page 262: ...Series 8500A Peak Power Meters Index 8 Manual No 20790 Rev C November 1998 ...