Reading and Defining Pulse Parameters
The following commands read the current values set for the pulse parameters:
* PUFT
Get Pulse Fall Time. The unit will respond with:
FALAddddddE-e (If channel A selected)
FALBddddddE-e (If channel B selected)
(d = seconds; e = exponent)
☛
NOTE: *PUFT, *PURT, and *PUWD operate only in the Graph Mode.
* PURT
Get Pulse Rise Time. The unit will respond with:
RISAddddddE-e (If channel A selected)
RISBddddddE-e (If channel B selected)
(d = seconds: e = exponent)
* PUWD
Get Pulse Width. The unit will respond with:
WIDAddddddE-e (If channel A selected)
WIDBddddddE-e (If channel B selected)
(d = seconds: e = exponent)
The following commands set new values for the pulse parameters:
PUFSddd
Set Pulse Fall Time - Start %.
Where d = % expressed in 0.1% units. Allowable range is 1 to 999 (0.1% to 99.9%)
PUFEddd
Set Pulse Fall Time - End %
Where d = % expressed in 0.1 % units. Allowable range is 1 to 999 (0.1% to 99.9%)
PURSddd
Set Pulse Rise Time - Start %
Where d = % expressed in 0.1% units. Allowable range is 1 to 999 (0.1% to 99.9%)
PUREddd
Set Pulse Rise Time - End %
Where d = % expressed in 0.1 % units. Allowable range is 1 to 999 (0.1% to 99.9%)
PUWSddd
Set Pulse Width - Start %
Where d = % expressed in 0.1 % units. Allowable range is 1 to 999 (0.1% to 99.9%)
PUWEddd
Set Pulse Width - End %
Where d = % expressed in 0.1% units. Allowable range is 1 to 999 (0.1% to 99.9%)
Remote Operation
Manual No. 20790, Rev C, November 1998
3-25
Summary of Contents for 8501A
Page 12: ...Series 8500A Peak Power Meters x Manual No 20790 Rev C November 1998 ...
Page 16: ...Series 8500A Peak Power Meters xiv Manual No 20790 Rev C November 1998 ...
Page 18: ...Series 8500A Peak Power Meters xvi Manual No 20790 Rev C November 1998 ...
Page 74: ...Series 8500A Peak Power Meters 2 48 Manual No 20790 Rev C November 1998 ...
Page 156: ...Series 8500A Peak Power Meters 4 30 Manual No 20790 Rev C November 1998 ...
Page 180: ...Series 8500A Peak Power Meters 5 24 Manual No 20790 Rev C November 1998 ...
Page 230: ...Series 8500A Peak Power Meters 8 2 Manual No 20790 Rev C November 1998 ...
Page 236: ...Series 8500A Peak Power Meters A 6 Manual No 20790 Rev C November 1998 ...
Page 262: ...Series 8500A Peak Power Meters Index 8 Manual No 20790 Rev C November 1998 ...