74
EN
Appendix
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer
and/or user of the device should assure that it is used in such an environment.
Emission test
Compatibility
Notes concerning the electromagnetic
environment
RF emissions CISPR 11
Group 1
RF energy is used only to maintain the device’s
operation. Therefore, its RF emissions are so
low that it is not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establish-
ments, other than domestic and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
Compatible
Guidance and manufacturer´s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
and/or user of the device should assure that it is used in such an environment.
Immunity test
IEC 60601 -
Test level
Compatibility
level
Notes concerning the electromag-
netic environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient / Burst
IEC 61000-4-4
Power supply
lines:
± 2 kV
Input/output lines:
± 1 kV
Power supply
lines:
± 2 kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge voltage
IEC 61000-4-5
Line(s) to line(s):
± 1 kV
Line(s) to earth/
ground:
± 2 kV
100 kHz repetition
frequency
Line(s) to line(s):
± 1 kV
100 kHz repetition
frequency
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % 0.5 cycle at
0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % 1 cycle and
70 % 25/30 cycles
Single-phase: at 0
0 % 300 cycles
0 % 0.5 cycle at
0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0% 1 cycle and
70 % 25/30
cycles
Single-phase: at 0
0 % 300 cycles
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency
(50 Hz / 60Hz)
Magnetic field
IEC 61000-4-8
30 A/m
50 Hz / 60 Hz
30 A/m
50 Hz / 60 Hz
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
Summary of Contents for Smart GT-1775
Page 2: ......
Page 3: ...DE Digitales Blutdruckmessger t f r den Oberarm Gebrauchsanweisung GT 1775...
Page 41: ...EN Digital Blood Pressure Monitor for the Upper Arm INSTRUCTION MANUAL GT 1775...
Page 79: ...FR Tensiom tre lectronique au bras NOTICE DE L UTILISATEUR GT 1775...
Page 117: ...IT Misuratore digitale della pressione arteriosa da braccio ISTRUZIONI D USO GT 1775...
Page 155: ...ES Tensi metro digital para brazo Manual de uso GT 1775...
Page 193: ...PT Esfigmoman metro digital de bra o MANUAL DE INSTRU ES GT 1775...
Page 231: ...NL Digitale bloeddrukmeter voor de bovenarm GEBRUIKSAANWIJZING GT 1775...
Page 269: ...AR AR AR GT 1775...
Page 272: ...272 AR 3...
Page 273: ...273 AR Geratherm Medical AG 22...
Page 274: ...274 AR 45 30 10 5 START STOP...
Page 275: ...275 AR 300 WLAN 4 9 10000...
Page 276: ...276 AR ISO ISO 10993 5 2009 10993 10 2010 APG AP 30...
Page 277: ...277 AR Geratherm smart Geratherm Medical AG Geratherm Medical AG Geratherm smart...
Page 278: ...278 AR WHO 80 120 84 80 129 120 89 85 139 130 99 90 159 140 1 109 100 179 160 2 110 180 3...
Page 279: ...279 AR 24 mmHg...
Page 280: ...280 AR mmHg mmHg min mmHg mmHg min START STOP LCD 1 2 3 4 5 Geratherm 42 22...
Page 281: ...281 AR mmHg mmHg min 0 133 1...
Page 282: ...282 AR 1 5 AAA 4...
Page 283: ...283 AR 24 2057 2017 START 1 3 STOP YEAR M 2...
Page 284: ...284 AR START STOP 3 YEAR DAY MONTH MONTH 3 2 4 DAY mmHg mmHg min...
Page 285: ...285 AR HOUR 3 2 5 MINUTE LCD 6 DONE...
Page 286: ...286 AR 1 2 2 3 cm 3 2...
Page 287: ...287 AR 1 6 5 5 3...
Page 288: ...288 AR START 1 STOP 0 LCD START STOP 2...
Page 289: ...289 AR 25 15...
Page 290: ...290 AR P 80 179 160 159 140 139 130 129 120 120 SYS 110 109 100 99 90 89 85 84 80 80 DIA...
Page 291: ...291 AR 1 M 3 M 2 6 00 1 2017 1...
Page 292: ...292 AR 1 3 M DEL ALL 2 START STOP START STOP M 3...
Page 293: ...293 AR 20...
Page 294: ...294 AR E 01 E 02...
Page 295: ...295 AR E 03 E 04 XX 02 01 EExx out...
Page 296: ...296 AR 4 1 5 AAA LCD 50 65 0 299 230 130 199 40 40 5 0 4 3 5 40 5 15 90 50 700 1060...
Page 297: ...297 AR 60 20 93 50 32 22 42 22 225 68 5 108 2 120 2 AAA 4 1 5 BF IP21 12 5 ME A04...
Page 301: ...301 AR BF YYMMXXX SN 93 0 93 0 SN 20 60 C 20 C 60...
Page 302: ...302 AR EMV EN 60601 1 2 EMV Geratherm EN 60601 1 2 Medical AG...
Page 303: ...303 AR CISPR 11 1 CISPR 11 IEC 61000 3 2 IEC 61000 3 3...
Page 307: ...307 AR UKW KW 80 150 3...
Page 310: ......
Page 311: ...Geratherm Medical AG Fahrenheitstr 1 Geschwenda 98716 Deutschland...