Table 1 Remote Control Commands
Note: These commands are common to all communication interfaces.
System
Command Description
Response
*IDN?
Queries the product number
*GP PDLE-101 V1.1#
*VER?
Queries the Firmware Version
*V1.0.100524A-G# (Sample response)
*SER?
Queries instrument serial number
Example: *
136000000007#
*ADR nn#
Set GPIB Address nn
Range: 1
≤
nn
≤
30
*E00# or other error code
*ADR?
Queries the GPIB ADDRESS
Response: *FF#
Example: *10# GPIB address set to 10
*WAV xxxx#
Set the Wavelength
Units: nm
Range: 1520 nm
≤
xxxx
≤
1570 nm for C band unit
*E00# or other error code
*WAV?
Queries wavelength setting.
Units: nm.
Response: *FFFF#
Example: *1550# wavelength setting 1550nm
*HOM#
Go to PDL control mode (main screen) without changing
PDL value
*E00# or other error code
*MOD?
Queries Current MODE Setting
(corresponds to front panel display)
Possible Responses:
*HOM# Main
screen
*OPT# Function/Settings
selection
*TRI DIS# Triangular Variation Disabled
*RAN DIS# Random Variation Disabled
*SQU DIS# Square Variation Disabled
*SIN DIS# Sine Variation Disabled
*TRI ENA# Triangular Variation Enabled
*RAN ENA# Random Variation Enabled
*SQU ENA# Square Variation Enabled
*SIN ENA# Sine Variation Enabled
*FUN OP1# Function selection option 1-triangle
*FUN OP2# Function selection option 2-random
*FUN OP3# Function selection option 3-sine
*FUN OP4# Function selection option 4-square
*SET OP1# Settings selection option 1- trigger
*SET OP2# Settings selection option 2-
λ
*SET OP3# Settings selection option 3- dyn. IP
*SET OP4# Settings selection option 4- static IP
*SET OP5#
Settings selection option 5- static IP set
*SET OP6# Settings selection option 6- GPIB
*ADD#
GPIB address selection
*WAV# wavelength
selection
*IPD#
IP Address: Dynamic Mode
*IPS#
IP Address: Static Mode
*TRG#
Trigger Setup Mode
Manual PDL Control
*PDL FF.f#
Set the static PDL Value FF.f
Range: 0.1 dB
≤
FF.f
≤
20 dB
*E00# or other error code
*PDL?
Queries current PDL setting.
Units: dB
Range: 0.1 dB
≤
FF.f
≤
20.0 dB
Response: *FF.f#
Example: *17.2# PDL set to 17.2 dB
Trigger Control
*TRG ENA#
Enable Trigger Mode
Note: Trigger pulse will only start function if PDLE is in
“Wait-For-Trigger Mode”
*E00# or other error code
Document #: GP-UM-PDLE-101-11
Page 27 of 61
Summary of Contents for PDLE-101
Page 31: ...3 Click Next 4 Click Next Document GP UM PDLE 101 11 Page 31 of 61 ...
Page 32: ...5 Check Specify a location then click Next Document GP UM PDLE 101 11 Page 32 of 61 ...
Page 36: ...12 Click Next 13 Click Finish Document GP UM PDLE 101 11 Page 36 of 61 ...
Page 38: ...In this example the USB device is connected to COM3 Document GP UM PDLE 101 11 Page 38 of 61 ...
Page 49: ... Properties on General tab Document GP UM PDLE 101 11 Page 49 of 61 ...