![GE Transport PT878 Abridged Manual Download Page 27](http://html1.mh-extra.com/html/ge/transport-pt878/transport-pt878_abridged-manual_4169150027.webp)
August 2008
2-4
Initial Setup
Special Transducers
(cont.)
5.
Use the [
T
] key to reach the
Wedge Tmp
prompt and press
[ENTER]
. Use the numeric keys to enter the wedge temperature
value provided with your transducer. Press
[ENTER]
to confirm
your entry.
6.
Use the [
T
] key to reach the
Wedge SS
prompt and press
[ENTER]
.
Use the numeric keys to enter the wedge sound speed (in ft/sec or
m/sec) value provided with your transducer. Press
[ENTER]
to
confirm your entry.
7.
Press
[F3]
(OK) to confirm your selections and exit the
Program
Menu
. Otherwise, you can press
[F2]
(Cancel) or the
[ESC]
key to
exit the
Program Menu
without confirming your selections.
Proceed to the next section to enter your pipe parameters.
Entering the Pipe
Parameters
To program your pipe parameters, scroll to the
Pipe
option on the
Program Menu
(see Figure 2-3 below) and press
[ENTER]
.
Note:
Refer to
Sound Speeds and Pipe Size Data (914-004)
for
additional information about pipe sizes and sound speeds.
Figure 2-3: The Pipe Option Window
1.
At the
Material
prompt, press
[ENTER]
to open a list of options (see
Table 3-1 on the next page). Use the [
T
] or [
S
] keys to scroll to
your material or choose
“Other”
if your material is not on the list.
Transducer
Lining
Path
Transducer/Pipe
Pipe
Fluid
Cancel
OK
Material
Other
Sound Speed
600.3 m/s
5
ANSI
OD, mm
Wall, mm
2
Nominal Schedule
OD x PI, mm
15.708
Measure Wall with TGauge
Summary of Contents for Transport PT878
Page 8: ...Chapter 1...
Page 22: ...Chapter 2...
Page 36: ...Chapter 3...
Page 61: ...Chapter 4...
Page 62: ...Managing Site Files The Site Menu 4 1 The Site Manager Window 4 2 The Site Manager Menu 4 3...
Page 75: ...Chapter 5...
Page 76: ...Data Logging The Logging Menu 5 1 The Log Manager Window 5 2 The Log Manager Menu 5 2...
Page 93: ...Appendix A...
Page 94: ...Menu Maps Main Menu A 1...