Chapter 2. Installation
14
PanaFlow™ MV80 & MV82 User’s Manual
2.2.3 Installing Flange-Style Flowmeters
Install the flange-style meter between two conventional pipe flanges of the same nominal size as the flowmeter (see
below). If the process fluid is a liquid, make sure the meter is located where the pipe is always full. This may
require locating the meter at a low point in the piping system.
Note:
Vortex flowmeters are not suitable for two-phase flows, such as liquid and gas mixtures.
For horizontal pipelines having a process temperature above 300° F, mount the meter at a 45 or 90-degree angle to
avoid overheating the electronics enclosure. To adjust the viewing angle of the enclosure or display/keypad, see
“Display/Keypad Adjustment (All Meters)” on page 30
“Enclosure Adjustment (Series MV80 Only)” on page 31
.
Figure 7: Flange-Style Flowmeter Installation
When installing the meter make sure the flange marked with a flow arrow is positioned upstream of the outlet flange,
with the arrow head pointing in the direction of flow. The mark is on the flange adjacent to the enclosure mounting
neck. This ensures that the sensor head is positioned downstream of the vortex shedder bar and is correctly aligned to
the flow. Installing the meter opposite this direction will result in completely inaccurate flow measurement.
CAUTION!
When using toxic or corrosive gases, purge the line with inert gas for a minimum of four
hours at full gas flow before installing the flowmeter.
FLOW
Enclosure and
display/keypad
are adjustable to suit
most viewing angles.
Incorrect gasket position!
Do not allow any gasket
material to extend into
the flow profile.
Shedder bar (bluff body)
is positioned upstream
of the sensor.
Summary of Contents for PanaFlow MV80
Page 2: ......
Page 12: ...Contents x PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 20: ...Chapter 1 Introduction 8 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 62: ...Chapter 2 Installation 50 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 148: ...Appendix B Approvals 136 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 158: ...Appendix D Glossary 146 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 160: ...Warranty 148 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 162: ... no content intended for this page ...
Page 163: ......