Chapter 5. Troubleshooting and Repair
5.1.2 Level Two Hidden Diagnostics Values (cont.)
•
Correction Pairs
•
ft3/sec
(1 through 10)
•
%Dev.
(1 through 10)
•
Roughness
= Factory use only.
•
Force Recal?
= Factory use only.
•
Min. Delta H
- Energy EMS meters only. Sets the deadband for totalization to begin. Must be greater than this
number (1 default) to initiate the totalizer.
•
Init Displ. (sec)
= Enter a value in seconds to initialize the display every xxx seconds. Enter a value of 0 to
disable initializing the display.
5.2 Analog Output Calibration
To check the 4-20 mA circuit, connect a DVM in series with the output loop. Select zero or full scale (from the second
column of the hidden diagnostics) and then press the
ENTER
key twice. This action will cause the meter to output its
4 mA or 20 mA signal. If the DVM indicates a current variation greater than ± 0.006 mA from the 4 mA or 20 mA
signal, adjust the setting up or down until the output is calibrated.
Note:
These settings are not for adjusting the output zero and span values to match a flow range. That function is
located in the Output Menu.
Summary of Contents for PanaFlow MV80
Page 2: ......
Page 12: ...Contents x PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 20: ...Chapter 1 Introduction 8 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 62: ...Chapter 2 Installation 50 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 148: ...Appendix B Approvals 136 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 158: ...Appendix D Glossary 146 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 160: ...Warranty 148 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 162: ... no content intended for this page ...
Page 163: ......