Chapter 4. Serial Communications
90
PanaFlow™ MV80 & MV82 User’s Manual
4.2.6g Exception Status Definitions
The Read Exception Status command (function code 07) returns the exception status byte, which is defined as shown in
below. This byte may be cleared by setting “coil” register #00003 (function code 5, address 2, data = 0xff00).
4.2.6h Discrete Input Definitions
The status of the three alarms may be monitored via the Modbus Read Discrete Input command (function code 02), as
shown in
below. The value returned indicates the state of the alarm, and will be 1 only if the alarm is enabled
and active. A zero value is transmitted for alarms that are either disabled or inactive.
4.2.6i
Control Register Definitions
The only writable registers in this implementation are the Reset Exception Status, Reset Meter and Reset Totalizer
functions, which are implemented as “coils” which may be written with the Write Single Coil command (function code
05) to address 8 through 10, respectively, (register #00009 through #00011). The value sent with this command must be
either 0x0000 or 0xff00, or the meter will respond with an error message; the totalizer will be reset or exception status
cleared only with a value of 0xff00.
Table 11: Exception Status BIts
Bit(s)
Definition
0-1
Byte order (see Modbus Order on
)
•
0 = 3-2:1-0 1 = 2-3:0-1
•
2 = 1-0:3-2 3 = 0-1:2-3
2
Temperature sensor fault
3
Pressure sensor fault
4
A/D converter fault
5
Period overflow
6
Pulse overflow
7
Configuration changed
Table 12: Alarm Status Registers
Registers
Variable
Function Code
Address
10001
Alarm #1 state
02
0
10002
Alarm #2 state
02
1
10003
Alarm #3 state
02
2
Summary of Contents for PanaFlow MV80
Page 2: ......
Page 12: ...Contents x PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 20: ...Chapter 1 Introduction 8 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 62: ...Chapter 2 Installation 50 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 148: ...Appendix B Approvals 136 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 158: ...Appendix D Glossary 146 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 160: ...Warranty 148 PanaFlow MV80 MV82 User s Manual no content intended for this page ...
Page 162: ... no content intended for this page ...
Page 163: ......