8-18
CIC Pro™
2026419-033E
Troubleshooting
WARNING
LOSS OF MONITORING — Provide alternate patient monitoring
or close observation before performing the system resource reset
procedure. Beds displayed on the CIC Pro center will not be
monitored during the system resource reset.
After system resource reset is complete and the monitoring function
at the CIC Pro center has been restored (about 30 seconds), verify
the correct monitoring state and alarm function.
8.
Restart the CIC Pro center.
See Perform safe restart of the CIC Pro center on
Activate licenses (manually) via the Activation Code Summary Sheet
Use this procedure when the license activation codes reside on the printed Activation
Code Summary Sheet.
The printed Activation Code Summary Sheet that matches the serial number of the
CIC Pro center you are the activating the options on is required.
1.
If you have not already logged onto Webmin,
2.
Click
Configuration > Licensing
.
3.
Next to the license you want to activate, type the activation code into the
Activation Code
field. See the Activation Code Summary Sheet.
4.
Click
Activate
to activate the license on this CIC Pro center.
NOTE
To remove an activated license, click
Remove
next to the license you want to
remove.
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......