2-20
CIC Pro™
2026419-033E
Equipment overview
If a bed goes offline, and another bed with the same name but a different IP address
comes online, it is interpreted as a new and different bed and a full disclosure data
store is created and utilized for it. However, the following conditions apply:
If the bed that went offline (with the original IP address) comes back online
within the
Offline Storage
setting time frame, the two beds continue to be treated
as separate, since the different IP addresses positively identify them as distinct
and separate.
If the bed that went offline (with the original IP address) does not come back
online within the
Offline Storage
setting time frame, the bed with the new IP
address is interpreted as a replacement for the bed with the original IP address. In
this case, the full disclosure data for the bed with the original IP address is
deleted, and full disclosure data collection continues, and the data is stored under
the bed with the new IP address only.
Unit Boundary
Each care unit operates independently with respect to storage of full disclosure
data. The only interaction is when displaying data across care units.
License mode: All beds within a care unit will have access to the same amount of
data storage (based on time) unless there is a failure to get a license (1 hour).
NOTE
This can be different across different care units (e.g., UnitA-72hours, UnitB-
None (1 hour).
Start-up mode: The same start-up rules apply to all beds within a care unit.
Default: The same full disclosure defaults are used on all CIC Pro centers within
a care unit.
Full disclosure master: One per care unit.
Acquisition/storage: The master CIC Pro center will only assign beds within the
care unit to CIC Pro centers within that care unit. At no time will data for a bed
be stored on a CIC Pro center in a different care unit.
Bed unit changes: If a bed changes its care unit name during full disclosure data
collection, all data for the bed is deleted. The care unit where the bed moved
becomes responsible for the bed.
CIC Pro center unit changes: If a CIC Pro center changes its care unit name
during full disclosure data collection, all data for all beds being having full
disclosure data collected by that CIC Pro center is deleted. The care unit with the
original name is responsible for the beds.
Networking
NOTE
The Unity Network has been renamed to the CARESCAPE Network. Not all
references to the Unity Network will be changed immediately; Unity may appear
in some places and CARESCAPE in others. It is important to understand that
while the CARESCAPE Network replaces the Unity Network name, they refer to
the same GE monitoring network.
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......