Configuration
Pre-configuration process
Pre-configuration requirements
All the CIC Pro centers connected to the CARESCAPE Network IX and MC
networks must comply with the following configuration requirements:
All devices must have the same time zone settings.
The
Automatically adjust clock for daylight saving changes
check box
must remain
UNCHECKED
at all times.
All MC network IP addresses must be in the same IP scheme with the same
subnet mask.
All IX network IP addresses must be in the same IP scheme with the same subnet
mask.
In the hierarchy of multiple compatible CIC Pro center hardware and software
versions that co-exist, it is important to assign a set of the highest MC network IP
addresses to the highest software version. For example, if you are installing CIC
v5.1.x to an existing CARESCAPE Network comprising of CIC v4.0.7 or later,
you must allocate a set of the highest MC network IP addresses to all of the CIC
Pro center’s hardware running CIC v5.1.x. No lower version CIC Pro center can
have an IP address on the MC network that is greater than the IP address of a
higher version CIC Pro center.
Use the Check Centrals utility to check for the following on all the CIC Pro centers on
the network:
Time zone settings
Daylight saving time (DST) status
CARESCAPE Network IP address errors
Status of 4.1.1-1 patch application on CIC Pro center v4.1.1 (available only with
the release of CIC Pro center v5.0.7 or later)
NOTE
The Check Centrals utility does
not
discover any version of Central Scope.
The Check Centrals utility
will
discover CIC Pro center hardware running
v1.5, but will not report its IP configuration information. This means that
you must physically locate these devices if they exist on the network. Please
read step
on page
The Check Centrals utility will
not
discover any non-CIC Pro center device
(e.g., printers, Aware Gateway, PDS) on the network.
The Check Centrals utility is also supplied separately. (For the part number of the
Check Centrals utility, see
Disaster recovery software kit and Service Tools CD on
.) You can run the utility directly from the CD on BCM and Nightshade
platforms. However, this utility must be copied from the CD to an NTFS formatted
USB memory stick for use on Bedrock platforms (v4.1.1 and greater).
Pre-configuration instructions
1.
Read and understand the following Caution and communicate this information to
the biomedical/clinical staff:
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......