Equipment overview
Licensing
All features and functions of the CIC Pro center are determined by the licenses
activated and running on each CIC Pro center. Licenses are specific to each individual
CIC Pro center’s serial number, are node locked, and cannot be used (floated) by
another CIC Pro center.
Full disclosure
The CIC Pro center full disclosure system stores all waveform and parametric data
from a patient for up to 72 hours. This data can be randomly accessed later in a static
display that looks similar to the real-time display window. To accomplish this, every
CARESCAPE Network waveform packet (4/sec) and every parameter packet (1
every 2 seconds) is stored on the CIC Pro center. In addition, one RWHAT packet and
one admit packet is stored every minute to help recreate the patient’s history.
Unlike an alarm history event stored at the bedside, which only stores a 10 second
snap shot surrounding the event, full disclosure allows the user to scroll back in time
(prior to the event and leading up to the event).
Behaviors or rules
Installation
Incompatible with previous full disclosure system (prior to CIC Pro center v3.x).
Certain
Unit Defaults
are incompatible with CIC Pro center v3.x systems.
Start-up modes
Licensing
Support for up to 16 patients per CIC Pro center.
Mode
Description
Auto For All
Automatically detects admitted beds that are not currently storing
full disclosure data anywhere and attempts to start full disclosure on
one of the CIC Pro centers within the care unit. This works
regardless of whether the bed was admitted directly at the bedside
or the CIC Pro center. Full disclosure cannot be stopped on a
particular bed until the bed is discharged.
Auto If Listed
Same as
Auto For All
, but only if the bed in question is entered into
the list on the
FD Unit Defaults
setup screen. All other beds are not
storing full disclosure data.
Manual Mode
Beds are not automatically storing full disclosure data upon
admission. Users can manually start and stop full disclosure for a
particular bed by using a button located on the
FD Strip
page in the
single viewer window. All full disclosure data is deleted when full
disclosure is stopped for a bed. See the CIC Pro Clinical
Information Center Operator’s Manual for more information.
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......