Configuration
2026419-033E
CIC Pro™
6-53
6.
To change an existing hardwire bed, telemetry bed, or telemetry transmitter/
transceiver, complete the following tasks:
a.
Use the scroll bar to locate the item you want to change.
b.
Click on the item to display an editing window.
c.
Press
Enter
to add the entry to the list.
7.
After making your selections, complete one of the following tasks:
Click
OK
to apply your changes and close the
CIC Setup
window.
Click
Apply
to apply your changes without closing the
CIC Setup
window.
8.
Verify the new settings prior to monitoring patients.
Set up locked beds
When logged into the Service mode, the CIC Pro center can be configured with the
bed names either in locked or unlocked mode. When locked, the bed name is
permanently allocated to a particular slot on the CIC Pro center and users are unable
to move the bed to another slot.
NOTE
It is possible to admit a patient to a window with a bed name that is locked to
NONE
. To avoid duplication of patient waveforms, a window locked as
NONE
should not be used to admit a patient.
Complete the following procedure to lock or unlock a bed:
1.
From the multi-patient viewer, click
Setup CIC
.
2.
Log on to Setup CIC with service access on page 4-5
.
3.
Position the mouse pointer in the patient’s waveform window and right-click.
4.
Click to choose either
LOCK
or
UNLOCK
. The change takes effect immediately.
(Chinese only) Set the pressures unit-of-measure
NOTE
This option is only available on CIC Pro centers running software v5.0.6 or later
configured for the Chinese language. Pressure values may be displayed in kPa
when configured for the Chinese language.
Complete the following procedure to set up the pressures values to display in kPa.
1.
On the
CIC Setup
screen, click
CIC Defaults
. The default pressures value is
mmHg
.
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......