6-72
CIC Pro™
2026419-033E
Configuration
4.
Right-click the
cic_xxxxxxxxx.cfd
link and select
Save Target As
to save the
backup file.
Planning your care unit defaults restore strategy
NOTE
Understand the functionality as described below before you proceed with the
restore procedure.
NOTE
The real-time trend graph setting is not saved during backup. It must be restored
manually.
When planning a restore strategy in a care unit running more than one CIC Pro center,
it is important to know that when a CIC Pro center starts up, it will update some of its
configuration data to match any other CIC Pro centers it finds on the care unit
network. The exact data that is automatically updated in this fashion is shown in
.
In general, this would be data common to the care unit that should be consistent
between all the CIC Pro centers within that care unit. All other configuration data is
specific to the local CIC Pro center only, and is not shared between the CIC Pro
centers in the care unit.
See the following scenarios for information about how this automatic update impacts
the backup and restoration process of a CIC Pro center.
Scenario 1: Restore a CIC Pro center in a care unit running only one CIC Pro
center on page 6-72
Scenario 3: Restore multiple CIC Pro centers in a care unit to a previous
configuration on page 6-74
Scenario 1: Restore a CIC Pro center in a care unit running only one CIC Pro
center
In this situation, automatic update does not occur, and is not a factor. All
configuration data written to the CIC Pro center during the restoration process
remains intact, since no other CIC Pro centers exist to trigger an automatic update.
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......