Installation
2026419-033E
CIC Pro™
5-7
Un-interruptible power supply (UPS)
WARNING
Connect the UPS to the CIC Pro center(s) and display monitor(s)
only. Do not connect printers or other devices to a UPS, as such
devices may shorten estimated run-times. If AC line power is not
restored before UPS run-time is exceeded, the CIC Pro center
improperly shuts down and patients will not be monitored.
WARNING
Without a UPS, power line outages may result in:
Improper shutdown of the CIC Pro center, causing lengthy disk
scan procedures on reboot.
Data loss.
If power to the CIC Pro center is lost, patient monitoring
information will no longer be displayed or stored.
GE recommends using a UPS with the CIC Pro center. See
for a listing of UPS units available for the CIC Pro center. Follow the manufacturer’s
recommendations for installing the UPS.
NOTE
The CIC Pro center draws approximately 40 watts of power. The displays each
draw approximately 50 watts of power. See the table on page
The following table identifies the estimated wattage for CIC Pro center MP100
systems and 19-inch medical displays:
The actual UPS run-times are variable, and are affected by the following conditions:
Battery age
Ambient temperature
Site-specific UPS usage patterns
Load characteristics
Number of CIC
Pro center
MP100
systems
Number of CIC Pro center
19-inch medical displays
Total load with CIC Pro center MP100
system(s) and CIC Pro center 19-inch
medical display(s) (estimated watts)
One
One display (primary)
75 W
One
Two displays (dual/
secondary)
115 W
Two
Four displays (Two primary,
Two secondary)
230 W
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......