4-16
MAC
TM
600
Revision G
2047426-002
Maintenance:
Preparing System for FRU Replacement
1. Open the device as described in
2. Raise the top cover of the housing and disconnect the keypad cable
from the J4 connector for the old PWA, or J7 connector for the new
PWA and display cable from the J5 connector for the old PWA or J6
connector for the new PWA. Disconnect the display cable connector
from the J2 connector on the Daughter PCB (refer
3. Remove the daughter board from the display holder as described in
“Replacing the Daughter PCB assembly”
.
4. Start with snap number 1, pull the snap out and push the display
holder in the direction of the arrow. Follow the same procedure for
snaps number 2, 3 and 4.
5. Once snaps number 1, 2, 3 and 4 are detached, lift the left side of the
display holder and remove the LCD Module from the cover top
pocket.
6. Pull out the display holder to the left to detach it from snaps number
5 and 6.
7. Make sure the O-ring is in place or replace it with the new one.
8. Replace the new LCD module in the pocket of cover top, with the
cable on the bottom side.
9. Place the display holder on the cover top with all the snaps aligned.
10. Press the display holder at each snap region, to assemble the display
holder to the cover top.
11. Snap the daughter board to the display holder.
12. Re-connect the display cable to J5 for the old PWA or J6 for the new
PWA and the keypad cable to J4 for the old PWA or J7 for the new
PWA.
13. Re-attach the upper section of the device to the lower section with the
four screws.
Summary of Contents for MAC 600 SPY
Page 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Page 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Page 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Page 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Page 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Page 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Page 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Page 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Page 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Page 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Page 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Page 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Page 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Page 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...