B-6
MAC
TM
600
Revision G
2047426-002
Electromagnetic Compatibility:
Guidance and Manufacturer’s Declaration-Electromagnetic Immunity
Recommended Separation Distance
The following table provides the recommended separation distances (in
meters) between portable and mobile RF communication equipment and
the MAC 600 system. The MAC 600 system is intended for use in the
electromagnetic environment on which radiated RF disturbances are
controlled. The customer or the user of the MAC 600 system can help
prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the MAC 600 system as recommended below,
according to the maximum output power of the communications
equipment.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (w)
according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Rated maximum output
power of transmitter in
watts (W)
Separation distance in meters (m) according to frequency of transmitter
150KHz to 80MHz
outside ISM bands
d =
1 . 2
150KHz to 80MHz outside
ISM bands
d =
1 . 2
80MHz to 800MHz
d
=
1 . 2
800MHz to 2.5GHz d
= 2.
3
0.01
0.12
0.12
0.12
0.23
0.1
0.38
0.38
0.38
0.73
1
1.2
1.2
1.2
2.3
10
3.8
3.8
3.8
7.3
100
12
12
12
23
Note 1: At 80MHz and 800MHz the separation distance of high frequency range applies.
P
P
P
P
Summary of Contents for MAC 600 SPY
Page 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Page 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Page 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Page 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Page 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Page 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Page 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Page 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Page 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Page 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Page 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Page 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Page 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Page 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...