Revision G
MAC
TM
600
4-29
2047426-002
Maintenance:
Check out Procedure
Verify the current leakage test results meet the requirements. Perform
electrical safety checks when indicated. All indicated electrical safety
checks require a pass/ fail indication for the steps performed. Record the
measurement values in your debrief.
1. Verify the equipment does not propose any health hazard. See
Leakage (AC line) Current Test”
on page 4-26, for more information.
2. Verify all exposed metal is properly grounded. See
on page 4-24 for more information.
NC = Normal Condition; SFC = Single Fault Condition; N/A = Not
Applicable
UUT = Unit Under Test
Table 5. Electrical Safety Checks
Step
Condition
UUT - ON
Result
Leakage Current Limits
Earth Leakage Current
1
Forward Polarity
NC
_______uA
Pass/Fail
150 uA
2
Neutral Open, Forward Polarity
SFC
_______uA
Pass/Fail
1000 uA
3
Neutral Open, Reverse Polarity
SFC
_______uA
Pass/Fail
1000
uA
4
Reverse Polarity
NC
_______uA
Pass/Fail
150
uA
Patient Leakage Current To Ground
1
Forward Polarity
NC
_______uA
Pass/Fail
10
uA
2
Neutral Open, Forward Polarity
SFC
_______uA
Pass/Fail
50 uA
3
Ground Open, Forward Polarity
SFC
_______uA
Pass/Fail
50 uA
4
Ground Open, Reverse Polarity
SFC
_______uA
Pass/Fail
50 uA
5
Neutral Open, Reverse Polarity
SFC
_______uA
Pass/Fail
50 uA
6
Reverse Polarity
NC
_______uA
Pass/Fail
10
uA
Ground Resistance
1
AC mains power cord ground prong
to exposed metal surface (ground lug)
N/A
_______
Q
Pass/Fail
Less than 1 Ohm
Summary of Contents for MAC 600 SPY
Page 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Page 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Page 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Page 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Page 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Page 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Page 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Page 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Page 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Page 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Page 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Page 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Page 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Page 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...