4-26
MAC
TM
600
Revision G
2047426-002
Maintenance:
Check out Procedure
Patient Leakage Current Tests
These tests measure leakage current of the device’s floating inputs
(patient connections ground).
Patient Leakage Current to Ground
This test measures leakage current from the floating input (patient
connection) to ground during normal condition (N.C.) and single fault
condition (S.F.C). In all cases, the leakage current is measured from the
floating input to ground: The device has to be turned on and off, and
connected to your safety testing instrument.
During N.C., measurements should be taken under the following
conditions (refer to
Device in ON state
Polarity switch NORM and RVS
GND switch GND closed
S1 (neutral) closed
During S.F.C., measurements should be taken under the following
conditions (refer to
Device in ON state
Polarity switch NORM and RVS
GND switch closed, S1 (neutral) open
GND switch open, S1 (neutral) closed
500 uA (IEC)
1,000 uA (IEC & UL)
150 uA (UL)
N.C.
S.F.C
Summary of Contents for MAC 600 SPY
Page 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Page 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Page 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Page 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Page 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Page 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Page 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Page 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Page 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Page 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Page 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Page 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Page 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Page 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...