Revision G
MAC
TM
600
3-9
2047426-002
Troubleshooting:
General Fault Isolation
Testing the Keypad
1. From the
Diagnostics
menu, select
Keyboard Test
.
The following window opens.
2. Press each key on the keypad and verify that the response appears in
the corresponding representation of that key on the screen.
When the corresponding key is pressed, a decimal value appears on
the screen. That will indicate whether or not the test has passed.
NOTE
Press
Esc
key in the end, as this will also exit from keyboard test.
See the following table for the decimal value matches that should be
present to pass the test.
Table 3-2. : Testing the Keypad
Key
Returned Decimal
Value
Returned
Character
Symbol on Display
Number key
1
49
1
1
Number key
2
50
2
2
Number key
3
51
3
3
Number key
4
52
4
4
Number key
5
53
5
5
Number key
6
54
6
6
Number key
7
55
7
7
Number key
8
56
8
8
Number
key
9
57
9
9
Number key
0
48
0
0
Key
Backspace
8
del
Function key
F1
1
F1
Summary of Contents for MAC 600 SPY
Page 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Page 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Page 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Page 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Page 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Page 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Page 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Page 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Page 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Page 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Page 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Page 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Page 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Page 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...