Revision G
MAC
TM
600
4-27
2047426-002
Maintenance:
Check out Procedure
The test has failed if the measured values are greater than:
Patient Leakage Current, Mains on Applied Part
WARNING
PERSONAL SAFETY — MAINS VOLTAGE is applied to
floating input (patient connection) and patient cables/test
plug conductors and safety tester PATIN JACK. For the
protection of the person performing these tests, the
following values of resistor R may be used:
Type CF — 100 kOhm (220 to 240 V)
Disconnect safety tester from line voltage before connecting
patient cable/test plug to device.
After line voltage has been disconnected, connect the
patient cable/test plug to device.
This test performs leakage current tests during single fault conditions
(S.F.C.) with line voltage applied to the floating (patient connection)
inputs.
N.C.
S.F.C
Polarity: NORM &
RVS S1
S1 (neutral): closed
GND: closed
Polarity: NORM &
RVS S1 (neutral): closed I
open GND: open I
Total tests/combinations: 2
Total tests/combinations: 4
10 uA (IEC & UL)
50 uA (EC & UL)
Summary of Contents for MAC 600 SPY
Page 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Page 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Page 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Page 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Page 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Page 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Page 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Page 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Page 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Page 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Page 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Page 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Page 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Page 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...